Product No: RB-3
RB-3 is designed according to the following standard requirements: IEC 60929 (GB/T15144), ANSI-ANSLG-C78.81, ANSI-ANSLG-C82.11, IEC62776, GB/T-15144 double-ended fluorescent lamp adoptting the IEC 60081 (GB/T10682) and the single-ended fluorescent lamp adoptting the IEC 60901 (GB/T17262)
The RB-3 works with HFP-800 (High Frequency Power Supply ) and LPS-3 (Auxiliary Power Supply) to be an IEC Standard Test System which used to test the Frequency Fluorescent Lamp such as T5 Tube. Please visit this page to view How to Measure the High Frequency Fluorescent Lamp.
Specifications:
• Available current range: 0~0.700A
• Impedance range: 100~2047Ω
• Impedance tolerance range: ±0.5% (Under calibration current). ±1% (Any current between 50% calibration current)
• Within the range of impedance: series inductance of impedance is less than 0.1mH, parallel connection capacitance is less than 1nF
LISUN RB-3 Adjustable Resistance Ballast – After Sales Questions and Answers.
LISUN RB-3 Adjustable Resistance Ballast were in the market for more than 10 years, LISUN engineers service team already summary the most of the After Sales Questions and Answers in the above link. Please read it carefuly, you can solve the most of the problems by yourself if you have some questions while using the instruments. LISUN will be continue to update this pages and more FAQ can be found in this topic link.