+8618117273997weixin
English
中文简体 中文简体 en English ru Русский es Español pt Português tr Türkçe ar العربية de Deutsch pl Polski it Italiano fr Français ko 한국어 th ไทย vi Tiếng Việt ja 日本語

Portable ESD Generator | Portable ESD Simulation Gun

Product No: ESD61000-2P

Leave a Message

=
  • Description
  • What is a ESD test?
    The static electricity generated by the human body to the object or between two objects may cause the electrical and electronic equipment circuits to malfunction or even be damaged. The Portable ESD Generator | Portable ESD Simulation Gun is designed for the withstanding ESD performance measurement for the assessment of electrical and electronic equipment. ESD61000-2P has a LCD touch screen in both English and Chinese. 

    What is an esd simulator gun used for?
    Portable ESD Generator | Portable ESD Simulation Gun is highest electrostatic voltage can be up to 20 kv, which is enough to cover the most severe standard grade of electrostatic voltage requirement (the voltage requirements of the level 4 air discharge electrostatic is 15 KV). The Portable ESD Generator | Portable ESD Simulation Gun can be used for the most of electrical and electronic equipment for electrostatic discharge test, and also can ensure comparability and reproducibility of test.

    Standard:
    IEC 61000-4-2/EN61000-4-2/GB/T17626.2 “Electromagnetic compatibility (EMC) – Part 4-2: Testing and measurement techniques – Electrostatic discharge immunity test”
    ISO10605 “Road Vehicles Test Methods for Electrical Disturbances from Electrostatic Discharge”
    GB/T17626.2 
    GB/T17215.322/IEC 62053-23:2003 “Electricity metering equipment (a.c.) – Particular requirements – Part 23: Static meters for reactive energy (classes 2 and 3)”

    Specification:

    Product Model ESD61000-2P
    Output Voltage 0.1~20kV±5%
    Polarity Positive/Negative
    Energy Storage Capacitance 150pF±10% (Replaceable)
    Discharge Resistor 330Ω±5% (Replaceable)
    Current Rise Time 0.6~1ns
    Testing Functions Single, Count, 20pps, Air, Contact, IEC level
    Trigger Mode MANUAL/AUTO
    Number of Discharge 1~9999
    Repetition 0.05s~99.99s
    Working Power AC110~240V, 50/60Hz

    Calibrate Report:

    Output Voltage (kV) First Peak Current (A) Current at 30ns (A) Current at 60ns (A) Rise Time (ns)
    2 7.29 4.1 2.2 0.93
    4 15.4 7.9 4.3 0.97
    6 23.2 12.1 6.5 0.97
    8 29.4 16.2 9.3 0.89
    -2 7.39 3.5 2.3 0.92
    -4 15.5 7.7 4.3 0.89
    -6 23.4 11.9 6.3 0.9
    -8 31.8 16.1 8.2 0.9