+8618117273997weixin
English
中文简体 中文简体 en English ru Русский es Español pt Português tr Türkçe ar العربية de Deutsch pl Polski it Italiano fr Français ko 한국어 th ไทย vi Tiếng Việt ja 日本語
09 Feb, 2023 822 Views Author: Cherry Shen

LISUN GLE-20 Face Masks Particulate Filtration Efficiency PFE Tester – After Sales Questions and Answers

GLE-20 is an oil-salt two-in-one Face Masks Particulate Filtration Efficiency PFE Tester. According to ASTM F1862-17ASTM F2299-03 (2017)ASTM F2100-2019BS EN14683-2019BS EN 14387-2004+A1 2008BS EN 136-1998BS EN 140-1999BS EN 143-2000 +C1-A1BS EN 149-2001+A1, etc. It is used to test the particulate filtering efficiency of daily protective masks and medical masks, and to determine the obstruction performance of ordinary fabrics and medical protective masks against a constant flow of airflow. It is widely used for medical device inspection centers, safety protection inspection centers, labor protection inspection centers, drug inspection centers, disease prevention and control centers, textile testing centers, hospitals, mask manufacturers, etc.

video

Standards:
1. It is designed and manufactured according to YY0469-2004 “Technical Requirements for Medical Surgical Masks” and GB2626-2006 “Respiratory Protective Equipment Self-Suction Filter Anti-Particle Respirator”. At the same time, it also refers to the advanced design concepts of similar international equipment (such as the American TSI company), based on the principle of “ European standard EN1822-3: 1998 single-sheet filter material test ”, but it is suitable for the test status of domestic related industries
2. GB / T 32610-2016 “Technical Specifications for Daily Protective Masks”
3.GB / T 19082-2009 “Technical requirements for medical disposable protective clothing”
4.GB / T 19083-2010 “Technical Requirements for Medical Protective Masks”
5.GB 24539-2009 General technical requirements for protective clothing and chemical protective clothing
6. YY / T 0969-2013 “Standard for disposable medical masks”

Q1: About air compressor
A1: The equipment needs to use air compressor to generate standard air flow through commissioning for filtration test. The equipment itself does not include an air compressor, which needs to be prepared by the customer.

Q2: The small ticket is not printed clearly
A2: Please open the cover. Check the ink cartridge and printer voltage display. If the ink cartridge is sufficient, the voltage needs to be fine-tuned to see the effect.

Q3: Use the same settings, but reduce the number of particles
A3: After continuous testing for about 24 hours, the number of particles may decrease. It is necessary to remove the container and clean the solution inside. If there is dirt, please also clean it. Then prepare new solution and pour it in.

LISUN GLE-20 Face Masks Particulate Filtration Efficiency PFE Tester – After Sales Questions and Answers

Container Location

Lisun Instruments Limited was found by LISUN GROUP in 2003. LISUN quality system has been strictly certified by ISO9001:2015. As a CIE Membership, LISUN products are designed based on CIE, IEC and other international or national standards. All products passed CE certificate and authenticated by the third party lab.

Our main products are GoniophotometerIntegrating SphereSpectroradiometerSurge GeneratorESD Simulator GunsEMI ReceiverEMC Test EquipmentElectrical Safety TesterEnvironmental ChamberTemperature ChamberClimate ChamberThermal ChamberSalt Spray TestDust Test ChamberWaterproof TestRoHS Test (EDXRF)Glow Wire Test and Needle Flame Test.

Please feel free to contact us if you need any support.
Tech Dep: Service@Lisungroup.com, Cell/WhatsApp:+8615317907381
Sales Dep: Sales@Lisungroup.com, Cell/WhatsApp:+8618117273997

Tags:

Leave a Message

Your email address will not be published. Required fields are marked *

=