Black Body Furnace | Tachometer Calibrator

Product No: BBF-1

Leave a Message

=
  • Description
  • Download
  • The BBF-1 Black Body Furnace is according to GB/T 21417.1-2008, GB/T21416-2008,etc. It is mainly used for the calibration of various electronic infrared ear thermometers and infrared forehead thermometers, and can also be used for the calibration of infrared body surface screening instruments.

    The blackbody furnace used for infrared thermometer calibration uses the temperature circulation system independently developed by our company, and the use of new materials to make the blackbody surface temperature more uniform. The emissivity can reach above 0.99, which can eliminate the error caused by the complex environment of the infrared temperature measurement system and effectively achieve a measurement accuracy of ±0.1℃. It can improve the reliability of the infrared temperature measurement system and reduce the error detection rate. The quality is reliable and has obtained the CNAS calibration certificate of a third-party laboratory.

    Specification:
    •  Operation temperature: 35 ℃ ~ 50 ℃ adjustable;
    •  Temperature resolution: 0.01 ℃;
    •  Temperature stability: ± (0.1 ~ 0.2) ℃ / 10min
    •  Cavity emissivity: ≥0.99; Cavity diameter: 55mm
    •  Heating time: More than 15 minutes
    •  Power supply: AC 110V~220V/50Hz, 100W
    •  Net Weight:3.5Kg; dimension:110mm*270mm*380mm(Height*Width*Depth)